- Products Details
Quick Details
-
Brand Name: Gz-lans
Model Number: LS-DH-P-42
Place of Origin: China
Packaging & Delivery
- Packaging Details: wooden box packing
- Delivery Detail: 3 moths
Specifications
Description:
The walk-in type VOCs Emission Test Room can simulate the temp,humidity,atmospheric pressure and airflow climate condition need by testing furniture,building materials,panel and some other light industry products.Regard variable flow clean air as fresh air supplement to the environment,ensure the accuracy of formaldehyde emission test.
Specification:
Item |
specifications |
||
Temp |
Test |
+20~+40℃ |
|
Temp range:(Zero load) |
40~80%RH(Temp within:+20~+40℃) |
||
Stability (Zero load) |
±0.5℃/±5% |
||
Uniformity (Zero load) |
±1℃/±5% |
||
Background concentration(+28℃/50%RH) (Zero load) |
Formaldehyde |
< 15mg/m³ |
|
TVOC |
< 20µg/m³ |
||
Single substance
|
< 2µg/m³ |
||
Air speed (Zero load) |
0.1~ 0.3m/s |
||
Ventilation rate(test process) |
0.2~ 2.0 1/h |
||
Ventilation precision |
≤±10% |
||
0.2 to 2.01/h |
|||
Degree of reduction |
≥80% (methybenzene) |
||
Inside demension |
W 4,000mm |
||
effective test space |
W 4,000mm |
||
internal capacity |
32m³ |
||
door size |
W 1,200 |
||
external dimension |
W 5,200mm |
||
spare parts |
No peper recorder, Sampling port (X2),sampling pump, cycling grading, lighting equipment,air compressor,water cleaner,Cable interface |
||
Installation demand |
|||
power supply |
Maximum capacity |
30kVA |
|
Power spec |
AC380V 3 phase 5 lines 90A |
||
air exhaust |
Airpipe size |
φ100 (φ97 ) |
|
Air flow |
Max. 72m3/h |
||
water supply |
Type |
tap water |
|
Flow |
maximun 65L/h |
||
water draining |
indirect water draining: DN50A |
||
floor carrying capacity |
250kg/m2 |
||
Total weight |
About 3,000kg |
Record parameters: temperature ,relative humidity,ventilation rate, atmospheric pressure, room, air , , , , , , pressure, cycle air flow.
Scale: 101 - 200 People
Country/Region: China (Mainland)/Asia
Established: 2018