This test laboratory is for measuring the capacity and power input of compressor condensing units, by secondary fluid calorimeter on the suction side (method A as per EN standard) for small unit or dry system refrigerant calorimeter on the suction side (Method B as per EN standard) for big unit and refrigerant liquid flow meter on the liquid side method (method D as per EN standard).
It consists of a climate chamber, a refrigerant test loop, an auxiliary water system and an electrical control and measurement system, to maintain compressor condensing unit in the specified working condition;
Feature at a glance
- Automatic control and stability monitoring for Suction temperature and pressure, Discharge pressure.
- Software application basing on Labview. Graphic user interface.
- secondary fluid calorimeter on the suction side or dry system refrigerant calorimeter on the suction side as primary method
- Refrigerant properties calculation traceable to NIST Database 23.
- Coriolis liquid mass flow meter method as secondary method
- Primary and secondary measurement, balance less than ±3%.
- The reproducibility less than ±1.5%.
- Capacity and power Difference compared with Master unit tested by the other test lab approved by certifying agencies≤ 2%
- Controlled temperature around the compressor is a standard specification.
Scale: 101 - 200 People